  "title": "SALA DE RAYOS X FIJA",
  "description": {
    "short": "Peligro o alteración del orden social, economía, servicios públicos, salubridad, seguridad o ambiente, por caso fortuito o de fuerza mayor.",
    "detailed": "SALA DE RAYOS X FIJA"
  },
  "lifecycle": {
    "section": "concluido",
    "source_status": "ADJUDICADO",
    "moved_at": null
  },
  "buyer": {
    "agency": "INSTITUTO NACIONAL DE ENFERMEDADES RESPIRATORIAS ISMAEL COSÍO VILLEGAS",
    "acronym": "INER",
    "branch": "12 - SALUD",
    "buying_unit": "DEPARTAMENTO DE ADQUISICIONES",
    "requesting_unit": "unidad requiriente data",
    "requesting_group": "DEPARTAMENTO DE ADQUISICIONES",
    "entity": "INSTITUTO NACIONAL DE ENFERMEDADES RESPIRATORIAS ISMAEL COSIO VILLEGAS",
    "slug": "iner",
    "unit_slug": "departamento-de-adquisiciones-012ncd001"
  },
  "classification": {
    "procedure_type": "ADJUDICACIÓN DIRECTA",
    "contracting_type": "ADQUISICIONES",
    "publication_type": null,
    "character": "NACIONAL",
    "fiscal_year": 2023,
    "law": "LEY DE ADQUISICIONES, ARRENDAMIENTOS Y SERVICIOS DEL SECTOR PÚBLICO",
    "legal_basis_id": "23",
    "exception_basis": "Art. 41 fr. II",
    "internal_control_number": "2023022017",
    "supply_source_number": null
  },
  "schedule": {
    "published_at": "2023-10-19T18:02:30.000Z",
    "clarification_meeting_at": null,
    "clarification_questions_deadline_at": null,
    "clarification_responses_deadline_at": null,
    "bid_opening_at": null,
    "award_at": null,
    "estimated_contract_at": "2023-10-27T06:00:00.000Z",
    "site_visit_at": null,
    "estimated_work_start_at": null
  },
  "estimated_value": {
    "minimum": null,
    "maximum": null,
    "currency": "MXN",
    "updated_at": null
  },
  "lots": [
    {
      "number": "1",
      "name": "PARTIDA SOLA",
      "description": "PARTIDA SOLA",
      "requirement_group_id": "1",
      "items": [
        {
          "number": "1",
          "cucop_code": "53101-1536",
          "specific_item_code": "53101",
          "cucop_description": "531.341.2511 UNIDAD RADIOLOGICA BASICA.. EQUIPO QUE PERMITE REALIZAR RADIOGRAFIAS DE TIPO GENERAL PARA CUBRIR LAS NECESIDADES PRIMARIAS DE UNA UNIDAD DE MEDICINA FAMILIAR DE 5 CONSULTORIOS. MESA FIJA: CON BUCKY INTEGRADO QUE ACEPTE CHASIS DE 35.56 X 43.18 CM. (14X17Š) CON REJILLA OSCILANTE FOCALIZADA A UN RANGO ENTRE 90 Y 105 CM. O MAYOR RELACION 8:1 O MAYOR ENTRE 30 Y 150 LINEAS POR CM. COLUMNA PORTATUBOS.: AUTOSOPORTABLE AL PISO CON LONGITUD DEL SOPORTE NO MAYOR DE 300 CM. CON DESPLAZAMIENTO SOBRE EL SOPORTE EN UN RANGO DE 225 A 275 CM. O MAYOR CON FRENOS ELECTROMAGNETICOS. TUBO DE RAYOS X: CON FOCO FINO DE 1.0 MM CON FOCO GRUESO DE 2.0 MM CON VELOCIDAD DEL ANODO DE 3000 RPM O MAYOR CON ROTACION DEL TUBO DE + 90 GRADOS. BUCKY VERTICAL: FIJO AL PISO O AL MURO CON REJILLA OSCILANTE CON RELACION 12: 1 CON FOCALIZACION DE 130 CM O MAYOR CON 30 LINEAS POR CM O MAYOR QUE ACEPTE CHASIS DE 35.56 X 43.18 CM. (14X17Š). GENERADOR: DE DOS PULSOS O DE ALTA FRECUENCIA CON CAPACIDAD DE UN RANGO DE 30 A 37.5 KW CON 300 MA O MAYOR CON UN RANGO DE 40 A 125 KV CON PASOS DE 2 KV CON RANGO DE TIEMPO DE 10 MILISEGUNDOS A 5 SEGUNDOS CON PANEL DE CONTROL DIGITAL QUE DESPLIEGUE: KV MA Y SEG.",
          "description": "\"SALA DE RAYOS X FIJA\n1. DESCRIPCIÓN\n1.- GENERADOR DE ALTA FRECUENCIA:\n1.1.- CON CAPACIDAD DE 80 KW. O MAYOR.\n1.2.- CON 500 MA. O MAYOR.\n1.3.-40 KV. A 150 KV. O MAYOR.\n1.4.- TIEMPO DE EXPOSICIÓN 1 MILISEGUNDO Ó MENOR A 5 SEGUNDOS.\n1.5.- CON PANEL DE CONTROL DIGITAL, QUE DESPLIEGUE: KV, MA Y SEG O MAS.\n2.- TUBO DE RAYOS X:\n2.1.- FOCO FINO DE 0.6 MM. O MENOR.\n2.2.- FOCO GRUESO DE 1.2 MM. MENOR.\n2.3.- ÁNODO ROTATORIO CON ÁNGULO DE 12°\n2.4.- CAPACIDAD DE ALMACENAMIENTO DE CALOR DEL ÁNODO 220 KJ O 300 KHU COMO MÍNIMO\n3.- COLUMNA PORTATUBOS:\n3.1.- SOPORTE A TECHO\n3.2.- CON DESPLAZAMIENTO LONGITUDINAL DE 180 CM. O MAYOR.\n3.4.- ANGULACIÓN DEL TUBO DE 0° Y ±120° MANUALES\n3.5.- CON FRENOS ELECTROMAGNÉTICOS.\n4.- DOS DETECTORES, UNO FIJO EN EL BUCKY Y UNO INALÁMBRICO(MÓVIL)\n4.1.- DETECTOR INALÁMBRICO QUE:\n4.2.- ADQUISICIÓN O PROFUNDIDAD DE IMAGEN DE 16 BITS O MAYOR O CONVERSIÓN ANALÓGICA-DIGITAL DE 14 BITS O MAYOR.\n4.3.- TAMAÑO DE MATRIZ DE IMAGEN DE 2300 X 2800 PIXELES COMO MÍNIMO\n4.4.- DE SELENIO AMORFO (A-SE) O SILICIO O SILICON AMORFO (A-SI) O IODURO DE CESIO.\n4.5.- TAMAÑO DE DETECTOR PORTABLE:\n4.5.1 TAMAÑO MÍNIMO DEL DETECTOR 35 CM X 43 CM (14\"\" X 17\"\")\n4.5.2 TAMAÑO MÍNIMO DEL ÁREA ACTIVA 34 CM X 43 CM (13.6\"\" X 16.6\"\") COMO MINIMO\n4.5.3.TAMAÑO DEL PIXEL DE 200 ¿M O MENOR.\n4.6.- DQE DE 50% O 3.1 IP./MM. O MAYOR.\n5.- TABLERO DE LA MESA CON RECORRIDO LONGITUDINAL ±20 CM. Ó MAYOR Y RECORRIDO LATERAL (TRANSVERSAL) ± 10 CM. Ó MAYOR.\n6.- CON ALTURA FIJA O ALTURA AJUSTABLE DE 60 CM. O MENOR A 80 CM. O MAYOR.\n7.- PORTADETECTOR CON REJILLA RELACIÓN 10:1 Ó MAYOR.\n8.- PORTADETECTOR PARA PARED O VERTICAL.\n9.- CONTROL AUTOMÁTICO DE EXPOSICIÓN.\n10.- CON SOPORTE DE PESO DEL PACIENTE DE 300 KG O MAYOR.\n11.- PROGRAMAS ANATÓMICOS O RADIOGRAFÍAS PROGRAMADAS O APR: 240 Ó MAYOR.\n12.- ESTACIÓN DE ADQUISICIÓN (INCLUIR MARCA Y MODELO CORRESPONDIENTE A LA MARCA DEL EQUIPO PRINCIPAL):\n12.1.- CON MONITOR DE 19\"\" O MAYOR Y RESOLUCIÓN DE 1K X 1K O MAYOR.\n12.2.- CAPACIDAD DE ALMACENAMIENTO DE 5000 IMÁG",
          "requested_quantity": "1",
          "minimum_quantity": "1",
          "maximum_quantity": null,
          "unit": "PIEZA",
          "minimum_amount": null,
          "maximum_amount": null
        }
      ]
    }
  ],
  "awards": [
    {
      "id": "C-2023-00130905",
      "internal_reference": "230496",
      "title": "SALA DE RAYOS X FIJA",
      "status": "PUBLICADO",
      "contractor": {
        "name": "SIEMENS HEALTHCARE DIAGNOSTICS S DE RL DE CV",
        "normalized_name": "SIEMENS HEALTHCARE DIAGNOSTICS",
        "rfc_type": null,
        "slug": "siemens-healthcare-diagnostics"
      },
      "contract_period": {
        "published_at": "2023-10-19T18:02:30.000Z",
        "starts_at": "2023-10-27T06:00:00.000Z",
        "ends_at": "2023-12-31T06:00:00.000Z"
      },
      "value": {
        "subtotal": "4716492.38",
        "tax": "754638.78",
        "total": "5471131.16",
        "maximum_subtotal": "0",
        "maximum_total": "0",
        "currency": "MXN"
      },
      "open_contract_type": null,
      "quantity": null,
      "exercised_quantity": null,
      "participants": [],
      "line_items": [
        {
          "number": "0",
          "cucop_code": "53101-1536",
          "description": "\"SALA DE RAYOS X FIJA\n1. DESCRIPCIÓN\n1.- GENERADOR DE ALTA FRECUENCIA:\n1.1.- CON CAPACIDAD DE 80 KW. O MAYOR.\n1.2.- CON 500 MA. O MAYOR.\n1.3.-40 KV. A 150 KV. O MAYOR.\n1.4.- TIEMPO DE EXPOSICIÓN 1 MILISEGUNDO Ó MENOR A 5 SEGUNDOS.\n1.5.- CON PANEL DE CONTROL DIGITAL, QUE DESPLIEGUE: KV, MA Y SEG O MAS.\n2.- TUBO DE RAYOS X:\n2.1.- FOCO FINO DE 0.6 MM. O MENOR.\n2.2.- FOCO GRUESO DE 1.2 MM. MENOR.\n2.3.- ÁNODO ROTATORIO CON ÁNGULO DE 12°\n2.4.- CAPACIDAD DE ALMACENAMIENTO DE CALOR DEL ÁNODO 220 KJ O 300 KHU COMO MÍNIMO\n3.- COLUMNA PORTATUBOS:\n3.1.- SOPORTE A TECHO\n3.2.- CON DESPLAZAMIENTO LONGITUDINAL DE 180 CM. O MAYOR.\n3.4.- ANGULACIÓN DEL TUBO DE 0° Y ±120° MANUALES\n3.5.- CON FRENOS ELECTROMAGNÉTICOS.\n4.- DOS DETECTORES, UNO FIJO EN EL BUCKY Y UNO INALÁMBRICO(MÓVIL)\n4.1.- DETECTOR INALÁMBRICO QUE:\n4.2.- ADQUISICIÓN O PROFUNDIDAD DE IMAGEN DE 16 BITS O MAYOR O CONVERSIÓN ANALÓGICA-DIGITAL DE 14 BITS O MAYOR.\n4.3.- TAMAÑO DE MATRIZ DE IMAGEN DE 2300 X 2800 PIXELES COMO MÍNIMO\n4.4.- DE SELENIO AMORFO (A-SE) O SILICIO O SILICON AMORFO (A-SI) O IODURO DE CESIO.\n4.5.- TAMAÑO DE DETECTOR PORTABLE:\n4.5.1 TAMAÑO MÍNIMO DEL DETECTOR 35 CM X 43 CM (14\"\" X 17\"\")\n4.5.2 TAMAÑO MÍNIMO DEL ÁREA ACTIVA 34 CM X 43 CM (13.6\"\" X 16.6\"\") COMO MINIMO\n4.5.3.TAMAÑO DEL PIXEL DE 200 ¿M O MENOR.\n4.6.- DQE DE 50% O 3.1 IP./MM. O MAYOR.\n5.- TABLERO DE LA MESA CON RECORRIDO LONGITUDINAL ±20 CM. Ó MAYOR Y RECORRIDO LATERAL (TRANSVERSAL) ± 10 CM. Ó MAYOR.\n6.- CON ALTURA FIJA O ALTURA AJUSTABLE DE 60 CM. O MENOR A 80 CM. O MAYOR.\n7.- PORTADETECTOR CON REJILLA RELACIÓN 10:1 Ó MAYOR.\n8.- PORTADETECTOR PARA PARED O VERTICAL.\n9.- CONTROL AUTOMÁTICO DE EXPOSICIÓN.\n10.- CON SOPORTE DE PESO DEL PACIENTE DE 300 KG O MAYOR.\n11.- PROGRAMAS ANATÓMICOS O RADIOGRAFÍAS PROGRAMADAS O APR: 240 Ó MAYOR.\n12.- ESTACIÓN DE ADQUISICIÓN (INCLUIR MARCA Y MODELO CORRESPONDIENTE A LA MARCA DEL EQUIPO PRINCIPAL):\n12.1.- CON MONITOR DE 19\"\" O MAYOR Y RESOLUCIÓN DE 1K X 1K O MAYOR.\n12.2.- CAPACIDAD DE ALMACENAMIENTO DE 5000 IMÁG",
          "unit": "PIEZA",
          "requested_quantity": "1",
          "awarded_quantity": null,
          "unit_price": "4716492.38",
          "subtotal": "4716492.38",
          "tax": "754638.78",
          "other_taxes": null,
          "total": "5471131.16"
        }
      ]
    }
  ],
  "documents": [
    {
      "id": null,
      "annex_number": 1,
      "document_type": null,
      "description": "JUSTIFICACIÓN DE EXCEPCIÓN",
      "filename": "just2017-000.pdf",
      "availability": "available",
      "size_bytes": "7078988",
      "mime_type": "application/pdf",
      "sha256": null,
      "uploaded_at": "2025-12-11T17:08:16.915Z",
      "updated_at": "2025-12-11T17:08:16.916Z"
    },
    {
      "id": null,
      "annex_number": 1,
      "document_type": null,
      "description": "ANEXO TÉCNICO",
      "filename": "ANEXO A 230496.pdf",
      "availability": "available",
      "size_bytes": "524719",
      "mime_type": "application/pdf",
      "sha256": null,
      "uploaded_at": "2025-12-11T17:08:19.536Z",
      "updated_at": "2025-12-11T17:08:19.538Z"
    },
    {
      "id": null,
      "annex_number": 4,
      "document_type": null,
      "description": "OFICIO DE ADJUDICACIÓN",
      "filename": "NOTIFICACION P-230496.pdf",
      "availability": "available",
      "size_bytes": "506970",
      "mime_type": "application/pdf",
      "sha256": null,
      "uploaded_at": "2025-12-11T17:08:21.056Z",
      "updated_at": "2025-12-11T17:08:21.057Z"
    }
  ],
  "alternates": [],
  "similar": [
    {
      "numero_procedimiento": "aa-86-x75-927006953-n-46-2023",
      "nombre_procedimiento": "EQUIPO MEDICO MEDIVAC",
      "dependencia": "SSALUDTAB",
      "siglas": "SSALUDTAB",
      "score": 0.09850568166408868
    },
    {
      "numero_procedimiento": "aa-50-gyr-050gyr040-t-15-2024",
      "nombre_procedimiento": "ADQUISICIÓN DE UNIDAD RADIOGRÁFICA DE 500MA, CUBIERTA DESPLAZABLE (DIGITAL)",
      "dependencia": "INSTITUTO MEXICANO DEL SEGURO SOCIAL",
      "siglas": "IMSS",
      "score": 0.10905075723235191
    },
    {
      "numero_procedimiento": "aa-84-w53-925002997-n-20-2025",
      "nombre_procedimiento": "ADQ.EQ.RAYOS X CEBPS SOL. CASGG",
      "dependencia": "084W53",
      "siglas": "084W53",
      "score": 0.137929268951967
    },
    {
      "numero_procedimiento": "aa-50-gyr-050gyr003-t-273-2023",
      "nombre_procedimiento": "BIENES DE INVERSION; UNIDAD RADIOLOGICA Y FLUOROSCOPICA TRANSPORTABLE TIPO ARCO",
      "dependencia": "IMSS",
      "siglas": "IMSS",
      "score": 0.15053437505969736
    },
    {
      "numero_procedimiento": "aa-50-gyr-050gyr029-t-466-2023",
      "nombre_procedimiento": "EQUIPAMIENTO UCIN HGZMF 3 MAZATLAN (ELECTRONICA Y MEDICINA)",
      "dependencia": "INSTITUTO MEXICANO DEL SEGURO SOCIAL",
      "siglas": "IMSS",
      "score": 0.1669797252088372
    },
    {
      "numero_procedimiento": "aa-50-gyr-050gyr075-t-94-2025",
      "nombre_procedimiento": "ARRENDAMIENTO DE EQUIPAMIENTO",
      "dependencia": "IMSS",
      "siglas": "IMSS",
      "score": 0.17174209187877576
    }
  ]
}